Calibrating odd-shaped or short sensors that will never fit a dry well, this calibration bath spans -30 to 150°C, holds ±0.1°C across the full range, and reaches 150°C in 30 minutes. The 90 x 90 mm opening takes any probe shape in up to 2.5 L of fluid. Auto step, ramp and soak run from the 6.5 in touch screen, with Wi-Fi and Ethernet for remote control.
Calibrating odd-shaped or short sensors that will never fit a dry well, this calibration bath spans -30 to 150°C, holds ±0.1°C across the full range, and reaches 150°C in 30 minutes. The 90 x 90 mm opening takes any probe shape in up to 2.5 L of fluid. Auto step, ramp and soak run from the 6.5 in touch screen, with Wi-Fi and Ethernet for remote control.
Features
Applications
| Temperature Performance | |
| Temperature Range at 23°C | -30 to 150°C |
| Display Accuracy | ±0.1°C |
| Stability | ±0.01°C (10 minutes) |
| Uniformity | ±0.02°C |
| Repeatability | 0.04°C internal reference, 0.01°C external reference |
| Resolution | Adjustable to 0.001°C |
| Temperature Units | °C, °F and K |
| Heating Time | 18 minutes from -30 to 23°C, 30 minutes from 23 to 150°C |
| Cooling Time | 45 minutes from 150 to 23°C, 45 minutes from 23 to -25°C, 55 minutes from 23 to -30°C |
| Typical Time to Stability | 10 minutes |
| Well and Insert | |
| Bath Opening | 90 x 90 mm |
| Maximum Media Depth | 6.10" (155 mm) |
| Bath Volume | 2.5 L maximum |
| Media Recommendation | 10 cSt silicone fluid |
| General | |
| Display | 6.5" (165 mm) color touch screen, 640 x 480 pixels |
| Communication | USB-A, USB-B, LAN, Wi-Fi, Bluetooth |
| Dimensions | 7.5 x 16.2 x 15.2" (191 x 412 x 387 mm) |
| Weight | 35.27 lb (16 kg) |
| Power Requirements | 100 to 120 V AC, 50/60 Hz, 8 A fuse |
| Environmental Conditions | 0 to 40°C operating, 8 to 33°C for guaranteed accuracy, 5% to 95% RH non-condensing, below 2000 m |
| Storage Conditions | -20 to 60°C |
| Mechanical Testing | Vibration 2 g; impact 5 g; package drop test 39.4" (1 m) |
| Temperature Performance | |
| Temperature Range at 23°C | -30 to 150°C |
| Display Accuracy | ±0.1°C |
| Stability | ±0.01°C (10 minutes) |
| Uniformity | ±0.02°C |
| Repeatability | 0.04°C internal reference, 0.01°C external reference |
| Resolution | Adjustable to 0.001°C |
| Temperature Units | °C, °F and K |
| Heating Time | 18 minutes from -30 to 23°C, 30 minutes from 23 to 150°C |
| Cooling Time | 45 minutes from 150 to 23°C, 45 minutes from 23 to -25°C, 55 minutes from 23 to -30°C |
| Typical Time to Stability | 10 minutes |
| Well and Insert | |
| Bath Opening | 90 x 90 mm |
| Maximum Media Depth | 6.10" (155 mm) |
| Bath Volume | 2.5 L maximum |
| Media Recommendation | 10 cSt silicone fluid |
| General | |
| Display | 6.5" (165 mm) color touch screen, 640 x 480 pixels |
| Communication | USB-A, USB-B, LAN, Wi-Fi, Bluetooth |
| Dimensions | 7.5 x 16.2 x 15.2" (191 x 412 x 387 mm) |
| Weight | 35.27 lb (16 kg) |
| Power Requirements | 100 to 120 V AC, 50/60 Hz, 8 A fuse |
| Environmental Conditions | 0 to 40°C operating, 8 to 33°C for guaranteed accuracy, 5% to 95% RH non-condensing, below 2000 m |
| Storage Conditions | -20 to 60°C |
| Mechanical Testing | Vibration 2 g; impact 5 g; package drop test 39.4" (1 m) |