Calibrating odd-shaped or short sensors that will never fit a dry well, this calibration bath spans -30 to 150°C, holds ±0.1°C across the full range, and reaches 150°C in 30 minutes. The 90 x 90 mm opening takes any probe shape in up to 2.5 L of fluid. Process electronics read a reference probe and two devices under test, drive HART transmitters and log results as found and as left.
Calibrating odd-shaped or short sensors that will never fit a dry well, this calibration bath spans -30 to 150°C, holds ±0.1°C across the full range, and reaches 150°C in 30 minutes. The 90 x 90 mm opening takes any probe shape in up to 2.5 L of fluid. Process electronics read a reference probe and two devices under test, drive HART transmitters and log results as found and as left.
Features
Applications
| Temperature Performance | |
| Temperature Range at 23°C | -30 to 150°C |
| Display Accuracy | ±0.1°C |
| Stability | ±0.01°C (10 minutes) |
| Uniformity | ±0.02°C |
| Repeatability | 0.04°C internal reference, 0.01°C external reference |
| Resolution | Adjustable to 0.001°C |
| Temperature Units | °C, °F and K |
| Heating Time | 18 minutes from -30 to 23°C, 30 minutes from 23 to 150°C |
| Cooling Time | 45 minutes from 150 to 23°C, 45 minutes from 23 to -25°C, 55 minutes from 23 to -30°C |
| Typical Time to Stability | 10 minutes |
| Well and Insert | |
| Bath Opening | 90 x 90 mm |
| Maximum Media Depth | 6.10" (155 mm) |
| Bath Volume | 2.5 L maximum |
| Media Recommendation | 10 cSt silicone fluid |
| Process Calibrator | |
| Reference Probe Input | 6-pin Lemo smart connector, 4-wire PRT |
| Reference Readout Accuracy | ±0.009°C at -40°C, ±0.017°C at 155°C, ±0.030°C at 420°C, ±0.042°C at 660°C |
| Reference Resistance Range | 0 to 400 Ω |
| Reference Resistance Accuracy | ±0.002 Ω from 0 to 50 Ω, ±40 ppm of reading from 50 to 400 Ω |
| Reference Characterizations | ITS-90, CVD, IEC-751, resistance |
| RTD Channels | 2, each accepting 2, 3 or 4-wire RTDs |
| RTD Measurement Accuracy | ±0.002 Ω from 0 to 25 Ω, ±0.008% of reading from 25 Ω to 4 kΩ |
| RTD Characterizations | PT10, PT25, PT50, PT100, PT200, PT500, PT1000, CU10, CU50, CU100, NI100, NI120 |
| Thermocouple Channels | 2, mini TC terminals accepting S, R, K, B, N, E, J, T, C, D, G, L and U |
| Thermocouple Accuracy, Type K | ±0.13°C at 0°C, ±0.16°C at 155°C, ±0.19°C at 350°C, ±0.25°C at 660°C |
| Thermocouple Range | -75 to 75 mV |
| Thermocouple Voltage Accuracy | 0.02% of reading + 5 µV |
| Internal Cold Junction Accuracy | ±0.35°C from 0 to 50°C ambient |
| Current Range | -30 to 30 mA |
| Current Accuracy | 0.02% of reading + 2 µA |
| Voltage Ranges | -12 to 12 V and -30 to 30 V |
| Voltage Accuracy | ±0.02% of reading + 2 mV |
| Switch Test | Mechanical or electrical |
| Loop Power Output | 24 V ±10%, 60 mA maximum |
| HART Communication | Read, configure and calibrate HART devices |
| Documentation | Up to 1000 tasks storing 10 results each, with as-found and as-left data |
| General | |
| Display | 6.5" (165 mm) color touch screen, 640 x 480 pixels |
| Communication | USB-A, USB-B, LAN, Wi-Fi, Bluetooth |
| Dimensions | 7.5 x 16.2 x 15.2" (191 x 412 x 387 mm) |
| Weight | 35.94 lb (16.3 kg) |
| Power Requirements | 100 to 120 V AC, 50/60 Hz, 8 A fuse |
| Environmental Conditions | 0 to 40°C operating, 8 to 33°C for guaranteed accuracy, 5% to 95% RH non-condensing, below 2000 m |
| Storage Conditions | -20 to 60°C |
| Mechanical Testing | Vibration 2 g; impact 5 g; package drop test 39.4" (1 m) |
Click on a category to view a selection of compatible accessories with the Additel ADT835PC-150-110V Portable Calibration Bath with Process Calibrator, -30 to 150°C, 110V.
| Temperature Performance | |
| Temperature Range at 23°C | -30 to 150°C |
| Display Accuracy | ±0.1°C |
| Stability | ±0.01°C (10 minutes) |
| Uniformity | ±0.02°C |
| Repeatability | 0.04°C internal reference, 0.01°C external reference |
| Resolution | Adjustable to 0.001°C |
| Temperature Units | °C, °F and K |
| Heating Time | 18 minutes from -30 to 23°C, 30 minutes from 23 to 150°C |
| Cooling Time | 45 minutes from 150 to 23°C, 45 minutes from 23 to -25°C, 55 minutes from 23 to -30°C |
| Typical Time to Stability | 10 minutes |
| Well and Insert | |
| Bath Opening | 90 x 90 mm |
| Maximum Media Depth | 6.10" (155 mm) |
| Bath Volume | 2.5 L maximum |
| Media Recommendation | 10 cSt silicone fluid |
| Process Calibrator | |
| Reference Probe Input | 6-pin Lemo smart connector, 4-wire PRT |
| Reference Readout Accuracy | ±0.009°C at -40°C, ±0.017°C at 155°C, ±0.030°C at 420°C, ±0.042°C at 660°C |
| Reference Resistance Range | 0 to 400 Ω |
| Reference Resistance Accuracy | ±0.002 Ω from 0 to 50 Ω, ±40 ppm of reading from 50 to 400 Ω |
| Reference Characterizations | ITS-90, CVD, IEC-751, resistance |
| RTD Channels | 2, each accepting 2, 3 or 4-wire RTDs |
| RTD Measurement Accuracy | ±0.002 Ω from 0 to 25 Ω, ±0.008% of reading from 25 Ω to 4 kΩ |
| RTD Characterizations | PT10, PT25, PT50, PT100, PT200, PT500, PT1000, CU10, CU50, CU100, NI100, NI120 |
| Thermocouple Channels | 2, mini TC terminals accepting S, R, K, B, N, E, J, T, C, D, G, L and U |
| Thermocouple Accuracy, Type K | ±0.13°C at 0°C, ±0.16°C at 155°C, ±0.19°C at 350°C, ±0.25°C at 660°C |
| Thermocouple Range | -75 to 75 mV |
| Thermocouple Voltage Accuracy | 0.02% of reading + 5 µV |
| Internal Cold Junction Accuracy | ±0.35°C from 0 to 50°C ambient |
| Current Range | -30 to 30 mA |
| Current Accuracy | 0.02% of reading + 2 µA |
| Voltage Ranges | -12 to 12 V and -30 to 30 V |
| Voltage Accuracy | ±0.02% of reading + 2 mV |
| Switch Test | Mechanical or electrical |
| Loop Power Output | 24 V ±10%, 60 mA maximum |
| HART Communication | Read, configure and calibrate HART devices |
| Documentation | Up to 1000 tasks storing 10 results each, with as-found and as-left data |
| General | |
| Display | 6.5" (165 mm) color touch screen, 640 x 480 pixels |
| Communication | USB-A, USB-B, LAN, Wi-Fi, Bluetooth |
| Dimensions | 7.5 x 16.2 x 15.2" (191 x 412 x 387 mm) |
| Weight | 35.94 lb (16.3 kg) |
| Power Requirements | 100 to 120 V AC, 50/60 Hz, 8 A fuse |
| Environmental Conditions | 0 to 40°C operating, 8 to 33°C for guaranteed accuracy, 5% to 95% RH non-condensing, below 2000 m |
| Storage Conditions | -20 to 60°C |
| Mechanical Testing | Vibration 2 g; impact 5 g; package drop test 39.4" (1 m) |