The latest in compact AC current probes. It is designed to meet the most stringent demands in industry and electrical contracting. This model meets EN 61010, 600V, Cat. III safety standards and is CE Marked and UL approved. With a small, compact size it features a measurement range of 0.5 to 240A AC.
The latest in compact AC current probes. It is designed to meet the most stringent demands in industry and electrical contracting. This model meets EN 61010, 600V, Cat. III safety standards and is CE Marked and UL approved. With a small, compact size it features a measurement range of 0.5 to 240A AC.
Discontinued!
This product has been discontinued and is no longer available.
There is no direct replacement, however we recommend the:
Not what you're looking for? Let’s do a search and help you find what you need.
The MN213 is a general purpose AC current probe, which is designed for DMMs and other measuring instruments with current input. It has a 1000:1 ratio and an output of 1mA per Amp in the primary (inside the jaw). The MN213 has integral 5 ft (1.5m) leads terminated by shrouded banana plugs. The MN213 has protective diodes across the output that limit the voltage in the case if the open secondary.
Features
Applications
| Nominal Range | 200A |
| Measurement Range | 0.5 to 240A |
| Transformation Ratio | 41.66736111111111 |
| Output Signal | 1mA/A on 1Ω |
| Accuracy (200A Range) | 0.5 to 10A: 3.0% of Reading ± 0.5A 10 to 40A: 2.5% of Reading ± 0.5A 40 to 100A: 2.0% of Reading ± 0.5A 100 to 240A: 1.0% of Reading ± 0.5A |
| Phase Shift (200A Range) | 0.5 to 10A: Not Specified 10 to 40A: ≤5.0° 40 to 100A ≤3.0° 100 to 240A: ≤2.5° |
| Overload | 240A for 10 min ON, 30 min OFF |
| Frequency Range | 40 to 10kHz |
| Limit Operating Conditions | 200A permanently to 1kHz Derating above 3kHz: 200A x (1/0.333F), F in kHz |
| Crest Factor | 3 at 200Arms with an error (due to CF) of 4% |
| Working/Common Mode Voltage | 600Vrms |
| Output Termination | 5ft Lead |
| Nominal Range | 200A |
| Measurement Range | 0.5 to 240A |
| Transformation Ratio | 41.66736111111111 |
| Output Signal | 1mA/A on 1Ω |
| Accuracy (200A Range) | 0.5 to 10A: 3.0% of Reading ± 0.5A 10 to 40A: 2.5% of Reading ± 0.5A 40 to 100A: 2.0% of Reading ± 0.5A 100 to 240A: 1.0% of Reading ± 0.5A |
| Phase Shift (200A Range) | 0.5 to 10A: Not Specified 10 to 40A: ≤5.0° 40 to 100A ≤3.0° 100 to 240A: ≤2.5° |
| Overload | 240A for 10 min ON, 30 min OFF |
| Frequency Range | 40 to 10kHz |
| Limit Operating Conditions | 200A permanently to 1kHz Derating above 3kHz: 200A x (1/0.333F), F in kHz |
| Crest Factor | 3 at 200Arms with an error (due to CF) of 4% |
| Working/Common Mode Voltage | 600Vrms |
| Output Termination | 5ft Lead |