| Technical Specifications |
| Display |
|
| |
VFD Dot Matrix |
| Resolution |
|
| |
6.5 Digits |
| V DC |
| Ranges | 100 mV to 1000 V | | Max. Resolution | 100 nV | | Accuracy | 8845A: 0.0035 + 0.0005 |
| 8846A: 0.0024 + 0.0005 | |
| Ranges |
100 mV to 1000 V |
| Max. Resolution |
100 nV |
| Accuracy |
8845A: 0.0035 + 0.0005 |
| |
8846A: 0.0024 + 0.0005 |
| V AC |
| Ranges | 8845A: 100 mV to 750 V |
| 8846A: 100 mV to 1000 V | | Max. Resolution | 100 nV | | Accuracy | 8845A: 0.06 + 0.03 |
| 8846A: 0.06 + 0.03 | | Frequency | 3 Hz to 300 KHz | |
| Ranges |
8845A: 100 mV to 750 V |
| |
8846A: 100 mV to 1000 V |
| Max. Resolution |
100 nV |
| Accuracy |
8845A: 0.06 + 0.03 |
| |
8846A: 0.06 + 0.03 |
| Frequency |
3 Hz to 300 KHz |
| Resistance |
| 2x4 Wire | Yes | | Ranges | 8845A: 100 Ω to 100 MΩ |
| 8846A: 10 Ω to 1 GΩ | | Max. Resolution | 8845A: 100 μΩ |
| 8846A: 10 μΩ | | Accuracy | 8845A: 0.010 + 0.001 |
| 8846A: 0.010 + 0.001 | |
| 2x4 Wire |
Yes |
| Ranges |
8845A: 100 Ω to 100 MΩ |
| |
8846A: 10 Ω to 1 GΩ |
| Max. Resolution |
8845A: 100 μΩ |
| |
8846A: 10 μΩ |
| Accuracy |
8845A: 0.010 + 0.001 |
| |
8846A: 0.010 + 0.001 |
| A DC |
| Ranges | 100 μA to 10 A | | Max. Resolution | 100 pA | | Accuracy | 0.050 + 0.005 | |
| Ranges |
100 μA to 10 A |
| Max. Resolution |
100 pA |
| Accuracy |
0.050 + 0.005 |
| A AC |
| Ranges | 8845A: 10 mA to 10 A |
| 8846A: 100 mA to 10 A | | Max. Resolution | 8845A: 10 nA |
| 8846A: 100 pA | | Accuracy | 0.10 + 0.04 | | Frequency | 3 Hz to 10 kHz | |
| Ranges |
8845A: 10 mA to 10 A |
| |
8846A: 100 mA to 10 A |
| Max. Resolution |
8845A: 10 nA |
| |
8846A: 100 pA |
| Accuracy |
0.10 + 0.04 |
| Frequency |
3 Hz to 10 kHz |
| Freq/Period |
| Ranges | 8845A: 3 Hz to 300 kHz |
| 8846A: 3 Hz to 1 MHz | | Max. Resolution | 1 μHz | | Accuracy | 0.01 % | |
| Ranges |
8845A: 3 Hz to 300 kHz |
| |
8846A: 3 Hz to 1 MHz |
| Max. Resolution |
1 μHz |
| Accuracy |
0.0001 |
| Continuity/ Diode Test |
|
| |
Yes |
| Capacitance |
| Ranges | 8846A: 1 nF to 0.1 F | | Max. Resolution | 8846A: 1 pf | | Accuracy | 8846A: 1 % | |
| Ranges |
8846A: 1 nF to 0.1 F |
| Max. Resolution |
8846A: 1 pf |
| Accuracy |
8846A: 1 % |
| Temperature |
| Type | 8846A: Platinum RTD | | Ranges | 8846A: -200°C to +600° | | Max. Resolution | 8846A: 0.01° | | Accuracy | 8846A: 0.06° | |
| Type |
8846A: Platinum RTD |
| Ranges |
8846A: -200°C to +600° |
| Max. Resolution |
8846A: 0.01° |
| Accuracy |
8846A: 0.06° |
| Math Functions |
| NULL, Min/Max/Average, Std Dev | | dB/dBm | Yes | |
| |
NULL, Min/Max/Average, Std Dev |
| dB/dBm |
Yes |
| Advanced Functions |
| Statistics/Histogram | Yes | | Trendplot | Yes | | Limit Test | Yes | |
| Statistics/Histogram |
Yes |
| Trendplot |
Yes |
| Limit Test |
Yes |
| Input Output |
| USB Memory | 8846A: USB memory drive port | | Real Time Clock | 8846A: Yes | | Interfaces | RS 232, IEE-488.2, Ethernet, USB (with optional adaptor) | | Programming Languages/Modes | 8846A: SCPI (IEEE-488.2), Agilent 34401A, Fluke 45 | |
| USB Memory |
8846A: USB memory drive port |
| Real Time Clock |
8846A: Yes |
| Interfaces |
RS 232, IEE-488.2, Ethernet, USB (with optional adaptor) |
| Programming Languages/Modes |
8846A: SCPI (IEEE-488.2), Agilent 34401A, Fluke 45 |
| General Specifications |
| Weight |
|
| |
3.6 kg (8.0 lbs) |
| Dimensions |
| 88 mm x 217 mm x 297 mm (3.46 in x 8.56 in x 11.7 in) |
|
| |
88 mm x 217 mm x 297 mm (3.46 in x 8.56 in x 11.7 in) |
| Safety |
| Designed to comply with IEC 61010-1 2000-1, ANSI/ISA-S82.01-1994, CAN/CSA-C22.2 No. 1010.1-92 CAT I 1000 V, CAT II 600 V |
|
| |
Designed to comply with IEC 61010-1 2000-1, ANSI/ISA-S82.01-1994, CAN/CSA-C22.2 No. 1010.1-92 CAT I 1000 V, CAT II 600 V |
| Warranty |
|
| |
One year |