Connect this probe to your Memory HiCorder or oscilloscope to capture select current signals from switching power supplies and more. The probe boasts a bandwidth of DC to 120MHz.
Connect this probe to your Memory HiCorder or oscilloscope to capture select current signals from switching power supplies and more. The probe boasts a bandwidth of DC to 120MHz.
Capture transient current signals from switching power supplies, inverters and motor controllers with this Hioki CT6701 Memory HiCorder Current Probe. When connected to your Memory HiCorder or oscilloscope's BNC terminal, the probe offers a wide bandwidth of DC to 120 MHz and features a 5 A continuous input.
Features
Frequency Bandwidth | DC to 120 MHz (-3 dB) |
Rise Time | 2.9 ns |
Noise Level | 60 μArms (typical), 75 μArms (maximum, for 30 MHz band-measuring instrument) |
Continuous Allowable Input | 5 Arms (DC and sine wave, requires derating at frequency) |
Maximum Allowable Peak Input | ±7.5 A peak (non-continuous) |
Amplitude Accuracy | ±1% of reading ±1 mV (DC, 45 to 66 Hz sine wave, 0 to 5 Arms, typical), ±3% of reading ±1 mV (DC, 45 to 66 Hz sine wave, 0 to 5 Arms, guaranteed) |
Output Rate | 1 V/A (when used with a device having a 1 MΩ input resistance or higher) |
Measurable Conductors | Insulated conductor |
Core Diameter | 0.2" (φ5 mm) |
Power Supply | ±12 V ±0.5 V, 3.2 VA |
Product Length | 1.02" (26 mm) sensor, 1.57" (40 mm) terminator, 4.92' (1.5 m) sensor cable with BNC terminal, 3.28' (1 m) power cable |
Product Width | 6.1" (155 mm) sensor, 1.14" (29 mm) terminator |
Product Height | 0.71" (18 mm) sensor, 3.27" (83 mm) terminator |
Product Weight | 8.8 oz (250 g) |
Frequency Bandwidth | DC to 120 MHz (-3 dB) |
Rise Time | 2.9 ns |
Noise Level | 60 μArms (typical), 75 μArms (maximum, for 30 MHz band-measuring instrument) |
Continuous Allowable Input | 5 Arms (DC and sine wave, requires derating at frequency) |
Maximum Allowable Peak Input | ±7.5 A peak (non-continuous) |
Amplitude Accuracy | ±1% of reading ±1 mV (DC, 45 to 66 Hz sine wave, 0 to 5 Arms, typical), ±3% of reading ±1 mV (DC, 45 to 66 Hz sine wave, 0 to 5 Arms, guaranteed) |
Output Rate | 1 V/A (when used with a device having a 1 MΩ input resistance or higher) |
Measurable Conductors | Insulated conductor |
Core Diameter | 0.2" (φ5 mm) |
Power Supply | ±12 V ±0.5 V, 3.2 VA |
Product Length | 1.02" (26 mm) sensor, 1.57" (40 mm) terminator, 4.92' (1.5 m) sensor cable with BNC terminal, 3.28' (1 m) power cable |
Product Width | 6.1" (155 mm) sensor, 1.14" (29 mm) terminator |
Product Height | 0.71" (18 mm) sensor, 3.27" (83 mm) terminator |
Product Weight | 8.8 oz (250 g) |