Employ the four-terminal DC method to quickly and accurately measure the resistance of components such as resistors and ferrite bead inductors. Includes advanced contact-check, comparator, and data export functions. The intuitive user interface and superb noise immunity are ideal for use with taping machines and separators.
Employ the four-terminal DC method to quickly and accurately measure the resistance of components such as resistors and ferrite bead inductors. Includes advanced contact-check, comparator, and data export functions. The intuitive user interface and superb noise immunity are ideal for use with taping machines and separators.
Features
With FAST measurement speed selected, measure resistance in as little as 0.9 ms*1 (including contact improvement, contact check and measurement) to decision output. Measure F-class (±1%) resistors at high speed. Use SLOW measurement speed to measure B-class (±0.1%) resistors in sync with the mains frequency. *1. In 100 or 1000 Ω measurement range, FAST speed, with low-power function disabled.
Comparator FunctionsCompare measurements against a specified reference value or range, with decision results available as signal outputs. User-friendly entry of comparator numerical values ensures smooth and reliable setting operations.
Store and Export MeasuredUp to 30,000 measurements can be stored in internal memory. Stored data can be exported to a computer as a batch, or used for statistical calculations.
Seven-Digit High-Resolution Display ("1,200,000")Perform high-resolution measurements on all E192-series resistance values, including B-class resistor testing.
Measurement types | Four-terminal resistance measurement 0.0000 mΩ (100 mΩ range) to 120.0000 MΩ Low-power four-terminal resistance measurement0.000 mΩ (1000 mΩ range) to 1200.000 Ω |
Measurement method | Four-terminal, constant-current DCMeasurement terminals: 22-mm BNC female jacks |
Range switching | Comparator on: Auto-range setting according to comparator reference or upper threshold setting. Comparator off: Manual range setting |
Zero-Adjust | Range: -1 to 10 Ω (wiring resistance compensation for two-terminal measurements) |
Trigger | Internal or External |
Sampling | Fast, Medium, and Slow |
Integration time setting function | 0.1 to 100.0 ms, PLC*2 setting available1 to 5 PLC @ 50 Hz, 1 to 6 PLC @60 Hz *2. One PLC = one power line cycle (mains wave-form period) |
Delay | DELAY1 = Set to allow for mechanical delay of trigger input and probing (affects all ranges), from 0.0 to 100.0 ms DELAY2 = Set to allow for measurement object response (each range independently), from 0.0 to 100.0 ms |
Functions | Self-calibration, probe short-circuit detection, Contact Improver, cur rent mode setting, OVC (offset voltage compensation), settings monitor, retry, statistical calculations, key-lock, comparator (relative tolerance or absolute range modes), EOM pulse width setting, data export, export data format, automemory |
Measurement fault detection functions | Out-of-range detection, contact check, current monitor, voltage monitor |
Memory storage | 30,000 values (volatile memory, no backup) |
Interfaces | EXT I/O, RS-232C, Printer, Settings MonitorFunctional terminals (SET MONITOR)GP-IB (Model RM5342-01) |
General Information | |
Dimensions | Approx. 260W × 88H × 300D mm (without projections) |
Weight | Approx. 2.9 kg |
Power consumption | 30 VA |
Rated mains supply frequency | 50 / 60 Hz |
Rated mains supply voltage | 100 to 240 V AC ±10% |
Measurement types | Four-terminal resistance measurement 0.0000 mΩ (100 mΩ range) to 120.0000 MΩ Low-power four-terminal resistance measurement0.000 mΩ (1000 mΩ range) to 1200.000 Ω |
Measurement method | Four-terminal, constant-current DCMeasurement terminals: 22-mm BNC female jacks |
Range switching | Comparator on: Auto-range setting according to comparator reference or upper threshold setting. Comparator off: Manual range setting |
Zero-Adjust | Range: -1 to 10 Ω (wiring resistance compensation for two-terminal measurements) |
Trigger | Internal or External |
Sampling | Fast, Medium, and Slow |
Integration time setting function | 0.1 to 100.0 ms, PLC*2 setting available1 to 5 PLC @ 50 Hz, 1 to 6 PLC @60 Hz *2. One PLC = one power line cycle (mains wave-form period) |
Delay | DELAY1 = Set to allow for mechanical delay of trigger input and probing (affects all ranges), from 0.0 to 100.0 ms DELAY2 = Set to allow for measurement object response (each range independently), from 0.0 to 100.0 ms |
Functions | Self-calibration, probe short-circuit detection, Contact Improver, cur rent mode setting, OVC (offset voltage compensation), settings monitor, retry, statistical calculations, key-lock, comparator (relative tolerance or absolute range modes), EOM pulse width setting, data export, export data format, automemory |
Measurement fault detection functions | Out-of-range detection, contact check, current monitor, voltage monitor |
Memory storage | 30,000 values (volatile memory, no backup) |
Interfaces | EXT I/O, RS-232C, Printer, Settings MonitorFunctional terminals (SET MONITOR)GP-IB (Model RM5342-01) |
General Information | |
Dimensions | Approx. 260W × 88H × 300D mm (without projections) |
Weight | Approx. 2.9 kg |
Power consumption | 30 VA |
Rated mains supply frequency | 50 / 60 Hz |
Rated mains supply voltage | 100 to 240 V AC ±10% |