Ideal for low-level drinking water, mid-range industrial, and high-range environmental applications. This meter meets the design criteria specified by ISO 7027 best for colored or extremely turbid samples. Also, it assures long-term stability and minimizes stray light and color interference.
Ideal for low-level drinking water, mid-range industrial, and high-range environmental applications. This meter meets the design criteria specified by ISO 7027 best for colored or extremely turbid samples. Also, it assures long-term stability and minimizes stray light and color interference.
Features
Applications
| Unit of Measure | |
| Ratiometric | FNRU, NTU, ASBC, EBC | 
| Nephelometric | FNU, NTU, ASBC, EBC | 
| Attenuation | FAU, NTU, ASBC, EBC | 
| Range | |
| Ratiometric | 0 to 1000 NTRU/FNRUC 0 to 17,500 ASBC 0 to 250 EBC | 
| Nephelometric | 0 to 100 NTRU/FNRUC 0 to 1750 ASBC 0 to 25 EBC | 
| Attenuation | 0 to 2000 AU/FAU 0 to 35,000 ASBC 0 to 500 EBC | 
| Resolution | |
| Ratiometric | 0 to 10.99 NTRU/FNRU: 0.01 11.0 to 109.9 NTRU/FNRU: 0.1 110 to 1000 NTRU/FNRU: 1 | 
| Nephelometric | 0 to 2.5 NTU/FNU: ±0.05 2.5 to 100 NTU/FNU: ±2% | 
| Attenuation | 0 to 2000 AU/FAU: ±10 AU/FAU or 6%, whichever is greater | 
| Detection Limit | |
| Ratiometric | 0.05 NTRU/FNRU | 
| Nephelometric | 0.05 NTU/FNU | 
| Attenuation | 10 AU/FAU | 
| Reproducibility | |
| Ratiometric | 0.02 NTRU/FNRU or 1% | 
| Nephelometric | 0.02 NTU/FNU or 1% | 
| Attenuation | 1% | 
| Range Selection | Automatic | 
| Light Source | IR LED 860 nm ±10 nm, spectral bandwidth with 50 nm | 
| Detector | Photodiode, centered at 90° and 180° | 
| Unit of Measure | |
| Ratiometric | FNRU, NTU, ASBC, EBC | 
| Nephelometric | FNU, NTU, ASBC, EBC | 
| Attenuation | FAU, NTU, ASBC, EBC | 
| Range | |
| Ratiometric | 0 to 1000 NTRU/FNRUC 0 to 17,500 ASBC 0 to 250 EBC | 
| Nephelometric | 0 to 100 NTRU/FNRUC 0 to 1750 ASBC 0 to 25 EBC | 
| Attenuation | 0 to 2000 AU/FAU 0 to 35,000 ASBC 0 to 500 EBC | 
| Resolution | |
| Ratiometric | 0 to 10.99 NTRU/FNRU: 0.01 11.0 to 109.9 NTRU/FNRU: 0.1 110 to 1000 NTRU/FNRU: 1 | 
| Nephelometric | 0 to 2.5 NTU/FNU: ±0.05 2.5 to 100 NTU/FNU: ±2% | 
| Attenuation | 0 to 2000 AU/FAU: ±10 AU/FAU or 6%, whichever is greater | 
| Detection Limit | |
| Ratiometric | 0.05 NTRU/FNRU | 
| Nephelometric | 0.05 NTU/FNU | 
| Attenuation | 10 AU/FAU | 
| Reproducibility | |
| Ratiometric | 0.02 NTRU/FNRU or 1% | 
| Nephelometric | 0.02 NTU/FNU or 1% | 
| Attenuation | 1% | 
| Range Selection | Automatic | 
| Light Source | IR LED 860 nm ±10 nm, spectral bandwidth with 50 nm | 
| Detector | Photodiode, centered at 90° and 180° |