The Advanced Pressure Calibrator is a general purpose, high accuracy pressure calibrator with special Features: intended to facilitate calibration of mass flow computers. With one internal pressure range and an available adapter for external BetaPort-P pressure modules, the possibilities are extensive. In addition, the advanced pressure calibrator has a port to connect a precision RTD probe for the temperature factor calibration required for electronic flow meters (EFMs).
The available internal sensor ranges (in PSI) are: 5, 15, 30, 100, 300, 500, 1000, 15000, 3000, 5000, and 10000. Pressure may be displayed in the user's choice of engineering units.
Ranges | |
RTD input range | (Pt100-385): -40 to 220°F (-40 to 105°C) |
Temperature Effect | No effect on accuracy on all functions from 69 to 95 °F (15 to 35 °C) Add
±0.002% F.S./°C for temps outside of 69 to 95 °F (15 to 35 °C) |
Engineering Units | psi, bar, mbar, kPa, kg/cm2, cmH2O@4°C, cmH2O@20°C, mmH2O@4°C, mmH2O@20°C, inH2O@4°C, inH2O@20°C, inH2O@60°F, mmHg@0°C, inHg@0°C, MPa, FT H2O@60°C plus 2 custom at time of order |
Accuracy | |
Pressure | 0.4 psi: ±0.1 % FS
1.0 psi: ±0.05 % FS 15 to 3000 psi: ±0.025 % FS 5,7, 5000, 10000 psi: ±0.035 % FS |
mA | ±0.015% of reading ±0.002mA |
Volts | ±0.015% of reading ±0.002V |
RTD input range (Pt100-385) | ±.01°C |
Temperature Effect | No effect on accuracy on all functions from 69 to 95 °F (15 to 35 °C)
Add ±0.002% F.S./°C for temps outside of 69 to 95 °F (15 to 35 °C) |
General | |
Operating Temperature | 14 to 122 °F (-10 to 50 °C) |
Storage Temperature | -4 to 140 °F (-20 to 60 °C) |
Power Supply | 4 x AA Batteries |
Ranges | |
RTD input range | (Pt100-385): -40 to 220°F (-40 to 105°C) |
Temperature Effect | No effect on accuracy on all functions from 69 to 95 °F (15 to 35 °C) Add
±0.002% F.S./°C for temps outside of 69 to 95 °F (15 to 35 °C) |
Engineering Units | psi, bar, mbar, kPa, kg/cm2, cmH2O@4°C, cmH2O@20°C, mmH2O@4°C, mmH2O@20°C, inH2O@4°C, inH2O@20°C, inH2O@60°F, mmHg@0°C, inHg@0°C, MPa, FT H2O@60°C plus 2 custom at time of order |
Accuracy | |
Pressure | 0.4 psi: ±0.1 % FS
1.0 psi: ±0.05 % FS 15 to 3000 psi: ±0.025 % FS 5,7, 5000, 10000 psi: ±0.035 % FS |
mA | ±0.015% of reading ±0.002mA |
Volts | ±0.015% of reading ±0.002V |
RTD input range (Pt100-385) | ±.01°C |
Temperature Effect | No effect on accuracy on all functions from 69 to 95 °F (15 to 35 °C)
Add ±0.002% F.S./°C for temps outside of 69 to 95 °F (15 to 35 °C) |
General | |
Operating Temperature | 14 to 122 °F (-10 to 50 °C) |
Storage Temperature | -4 to 140 °F (-20 to 60 °C) |
Power Supply | 4 x AA Batteries |