Designed for water treatment and purification systems and wastewater quality measurement, this conductivity electrode provides 10 to 10000 μS.
Designed for water treatment and purification systems and wastewater quality measurement, this conductivity electrode provides 10 to 10000 μS.
Conductivity cells for the CDCN-91, CDTX-90 and CDCN-5800 Series. The cells include a reversible fitting assembly for submersion mounting and use Pt1000 temperature compensation. Available cell constants are 0.01, 0.1, 1.0, 10 and 20.
Features
Applications
| Series | CDCE90000 Series |
| Type | conductivity electrode |
| Measurement range | 10 μS to 10000 μS |
| Temperature Compensation | Pt1000 |
| O-Rings | EPR |
| Insulator Material | PTFE |
| Electrodes | 316 SS |
| Standard Fitting | Polypropylene fitting 3/4" (19 mm) NPT thread |
| Maximum Pressure | 100 psi (6.9 bar) |
| Maximum Temperature | 212°F (100°C) |
| Optional Fitting | 316 SS 1/2" (12.7 mm) NPT |
| Optional Fitting Maximum Pressure | 200 psi (13.8 bar) |
| Optional Fitting Maximum Temperature | 248°F (120°C) |
| Cable | 15' (4.6 m) cable (standard) |
| Cell Constant / Conductivity Range (CDCE-90-001) | Cell constant 0.01; 0.010 to 100 μS (10 KΩ to 100 MΩ) |
| Cell Constant / Conductivity Range (CDCE-90-01) | Cell constant 0.1; 1 to 1000 μS |
| Cell Constant / Conductivity Range (CDCE-90-1) | Cell constant 1.0; 10 to 10,000 μS |
| Cell Constant / Conductivity Range (CDCE-90-10) | Cell constant 10.0; 100 to 200,000 μS; Insulator Material CPVC; Fitting Material 316 SS; Maximum Pressure/Temperature 100 psig (6.89 bar) @ 203°F (95°C) |
| Cell Constant / Conductivity Range (CDCE-90-20) | Cell constant 20.0; 200 to 400,000 μS; Insulator Material PTFE; Fitting Material 316 SS; Maximum Pressure/Temperature 100 psig (6.89 bar) @ 302°F (150°C) |
| Sanitary Models (CDCE-90S) | Tri-Grip Sanitary Fitting size 1, 1 1/2, 2" (12.7 mm, 5.1 cm); Electrodes 316 SS or titanium; Tri-Grip Sanitary Fitting 316 SS or titanium; Insulator Material PTFE; O-Ring EPR; Maximum Pressure 100 psi (6.9 bar); Maximum Temperature 248°F (120°C) |
| Series | CDCE90000 Series |
| Type | conductivity electrode |
| Measurement range | 10 μS to 10000 μS |
| Temperature Compensation | Pt1000 |
| O-Rings | EPR |
| Insulator Material | PTFE |
| Electrodes | 316 SS |
| Standard Fitting | Polypropylene fitting 3/4" (19 mm) NPT thread |
| Maximum Pressure | 100 psi (6.9 bar) |
| Maximum Temperature | 212°F (100°C) |
| Optional Fitting | 316 SS 1/2" (12.7 mm) NPT |
| Optional Fitting Maximum Pressure | 200 psi (13.8 bar) |
| Optional Fitting Maximum Temperature | 248°F (120°C) |
| Cable | 15' (4.6 m) cable (standard) |
| Cell Constant / Conductivity Range (CDCE-90-001) | Cell constant 0.01; 0.010 to 100 μS (10 KΩ to 100 MΩ) |
| Cell Constant / Conductivity Range (CDCE-90-01) | Cell constant 0.1; 1 to 1000 μS |
| Cell Constant / Conductivity Range (CDCE-90-1) | Cell constant 1.0; 10 to 10,000 μS |
| Cell Constant / Conductivity Range (CDCE-90-10) | Cell constant 10.0; 100 to 200,000 μS; Insulator Material CPVC; Fitting Material 316 SS; Maximum Pressure/Temperature 100 psig (6.89 bar) @ 203°F (95°C) |
| Cell Constant / Conductivity Range (CDCE-90-20) | Cell constant 20.0; 200 to 400,000 μS; Insulator Material PTFE; Fitting Material 316 SS; Maximum Pressure/Temperature 100 psig (6.89 bar) @ 302°F (150°C) |
| Sanitary Models (CDCE-90S) | Tri-Grip Sanitary Fitting size 1, 1 1/2, 2" (12.7 mm, 5.1 cm); Electrodes 316 SS or titanium; Tri-Grip Sanitary Fitting 316 SS or titanium; Insulator Material PTFE; O-Ring EPR; Maximum Pressure 100 psi (6.9 bar); Maximum Temperature 248°F (120°C) |