Designed for bringing analog process signals into an IIoT network, this process monitoring smart probe accepts 4 to 20 mA, 0-1 V, and 0 to 2 V inputs across 3 channels through an M12 5-pin connector. It includes 2 discrete I/O pins for alarms and control plus 10,000-point data logging. It connects to the Omega Link Cloud through a wired or wireless Smart Interface.
Designed for bringing analog process signals into an IIoT network, this process monitoring smart probe accepts 4 to 20 mA, 0-1 V, and 0 to 2 V inputs across 3 channels through an M12 5-pin connector. It includes 2 discrete I/O pins for alarms and control plus 10,000-point data logging. It connects to the Omega Link Cloud through a wired or wireless Smart Interface.
The Omega Link SP-014 Process Monitoring Smart Probe provides an easy way to integrate process signals to your Omega Link Ecosystem. The SP-014 accepts standard process signals through its M12 5-pin connector and Omega Link Smart Interfaces through its M12 8-pin connector. The optional M12.8-T-SPLIT Sensor Splitter can be used to access the Discrete I/O pins on the M12 8-pin connector. The optional M12.5-S-M-FM and M12.8-S-M-FM mating connectors can be utilized to easily connect wire leads to the SP-014 or Sensor Splitter.
The Omega Link SP-014 features 2 discrete I/O pins. These can be used for a myriad of applications including driving relays, physical alarms, or sensing dry contacts like door switches. The SP-014 can also be utilized as an edge controller, with autonomous independent decision-making capabilities to generate local alarms or provide control outputs based on sensor inputs.
Each of the 3 process inputs may be independently configured as 0-24 mA, 0-1.0 V DC, or 0-2.0 V DC inputs. A mixed-mode configuration option allows for the measurement of one process input and one digital pulse input which supports frequency (rate), pulse width, pulse duty cycle, and pulse counting (totalizer).
Features
Applications
| Voltage | 2.8 V DC - 3.3 V DC |
| ON | 1.0 V DC |
| OFF | 0.7 V DC |
| Internal Pull Up/Down | 1.5k to 3.0 V DC |
| Comparator (Pulse) Input | 100 mV, 500 mV, 1.0 V DC, 2.0 V DC |
| Frequency (Rate) | 0.01 Hz to 100 Hz, T PW MIN = 200 uS, accuracy ±0.5% |
| Frequency (Rate) | 100 to 1000 Hz, T PW MIN = 200 uS, accuracy ±1 Hz averaged over 1 sec. |
| Counter | 0 to +8388608, 1 kHz Maximum Rate, accuracy ±1 Count Maximum |
| Pulse Width (T PW) | 200 uS minimum, accuracy ±50 uS ±1% |
| Duty Cycle | 1% to 99%, 0.01 Hz to 1000 Hz, T PW MIN = 200 uS, accuracy ±1.5% Maximum |
| Current Loop | 0 - 24 mA, resolution ±0.1 mA, accuracy ±0.2 mA, input impedance 50 ohm |
| Voltage 0 - 1.0 V DC | resolution ±10 mV, maximum 1.20 V DC, accuracy ±10 mV, input impedance 100k ohm |
| Voltage 0 - 2.0 V DC | resolution ±10 mV, maximum 2.50 V DC, accuracy ±20 mV, input impedance 100k ohm |
| V inHighThreshold | 2.2 V maximum |
| V inLowThreshold | 0.3 V minimum |
| V inMAX | 30 V DC |
| Type | 2x Open Drain 100 mA maximum |
| V MAX | 30 V DC |
| Operating temperature | -40 to 1562°F (-40 to 850°C) |
| Rating | IP67 when mated |
| Agency approvals | CE, UKCA |
| Compatibility | Compatible with OEG, SYNC configuration software, Omega Link Cloud, and Modbus Networks |
| Dimensions | Dimensions: 0.87 x 3.80" (22.1 x 96.7 mm) not including mounting tabs |
| Voltage | 2.8 V DC - 3.3 V DC |
| ON | 1.0 V DC |
| OFF | 0.7 V DC |
| Internal Pull Up/Down | 1.5k to 3.0 V DC |
| Comparator (Pulse) Input | 100 mV, 500 mV, 1.0 V DC, 2.0 V DC |
| Frequency (Rate) | 0.01 Hz to 100 Hz, T PW MIN = 200 uS, accuracy ±0.5% |
| Frequency (Rate) | 100 to 1000 Hz, T PW MIN = 200 uS, accuracy ±1 Hz averaged over 1 sec. |
| Counter | 0 to +8388608, 1 kHz Maximum Rate, accuracy ±1 Count Maximum |
| Pulse Width (T PW) | 200 uS minimum, accuracy ±50 uS ±1% |
| Duty Cycle | 1% to 99%, 0.01 Hz to 1000 Hz, T PW MIN = 200 uS, accuracy ±1.5% Maximum |
| Current Loop | 0 - 24 mA, resolution ±0.1 mA, accuracy ±0.2 mA, input impedance 50 ohm |
| Voltage 0 - 1.0 V DC | resolution ±10 mV, maximum 1.20 V DC, accuracy ±10 mV, input impedance 100k ohm |
| Voltage 0 - 2.0 V DC | resolution ±10 mV, maximum 2.50 V DC, accuracy ±20 mV, input impedance 100k ohm |
| V inHighThreshold | 2.2 V maximum |
| V inLowThreshold | 0.3 V minimum |
| V inMAX | 30 V DC |
| Type | 2x Open Drain 100 mA maximum |
| V MAX | 30 V DC |
| Operating temperature | -40 to 1562°F (-40 to 850°C) |
| Rating | IP67 when mated |
| Agency approvals | CE, UKCA |
| Compatibility | Compatible with OEG, SYNC configuration software, Omega Link Cloud, and Modbus Networks |
| Dimensions | Dimensions: 0.87 x 3.80" (22.1 x 96.7 mm) not including mounting tabs |