Dual LCD display with analog bargraph indicates relative humidity and ambient temperature. Provides dew point and wet bulb readings.
Dual LCD display with analog bargraph indicates relative humidity and ambient temperature. Provides dew point and wet bulb readings.
Discontinued!
This product has been discontinued and is no longer available.
There is no direct replacement, however we recommend the:
Not what you're looking for? Let’s do a search and help you find what you need.
Features
Applications
Ideal for the following applications: Environmental Studies, HVAC, Coating Industry, Laboratory, Education, Musical Instruments, Food Distribution
Measuring Range | Temperature: -20 to 60°C(-4 to 140°F) Humidity: 10% RH to 95% RH |
Resolution | 0.1°C/0.1°F/0.1% RH |
Overrange Display | Highest digit of (1) is displayed |
Accuracy | Temperature: ±0.5°C, ±0.9°F Humidity: ±3% RH (at 25°C, 30 to 95% RH); ±5% RH(at 25°C, 10 to 30% RH) |
Datalogging | Manual Recording: 99 data sets (Direct reading from LCD display) Continuous Recording: 15000 data sets (Automatic timer recording, RS-232 interface/software) |
Input Protection | 60V DC or 24 Vrms AC maximum input voltage on any Combination of input Connector |
Sampling Rate | 1 times/sec |
Measuring Rate | Approximately 2.5 time/sec |
Operating Conditions | 0 to 60°C (32 to 140°F) & 10 to 80% RH |
Operating and Storage Conditions | 0 to 50°C < 80% RH -10 to 60°C < 70% RH |
Power Source | 6 x AAA Size Battery |
Battery Life (typical) | 200 hours |
Size & Weight | 152(L) ×72(W) ×35(H)mm & Approx. 235g |
Measuring Range | Temperature: -20 to 60°C(-4 to 140°F) Humidity: 10% RH to 95% RH |
Resolution | 0.1°C/0.1°F/0.1% RH |
Overrange Display | Highest digit of (1) is displayed |
Accuracy | Temperature: ±0.5°C, ±0.9°F Humidity: ±3% RH (at 25°C, 30 to 95% RH); ±5% RH(at 25°C, 10 to 30% RH) |
Datalogging | Manual Recording: 99 data sets (Direct reading from LCD display) Continuous Recording: 15000 data sets (Automatic timer recording, RS-232 interface/software) |
Input Protection | 60V DC or 24 Vrms AC maximum input voltage on any Combination of input Connector |
Sampling Rate | 1 times/sec |
Measuring Rate | Approximately 2.5 time/sec |
Operating Conditions | 0 to 60°C (32 to 140°F) & 10 to 80% RH |
Operating and Storage Conditions | 0 to 50°C < 80% RH -10 to 60°C < 70% RH |
Power Source | 6 x AAA Size Battery |
Battery Life (typical) | 200 hours |
Size & Weight | 152(L) ×72(W) ×35(H)mm & Approx. 235g |